Gf-signet 2250 Hydrostatic Level Sensor Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Nástroje Gf-signet 2250 Hydrostatic Level Sensor. GF Signet 2250 Hydrostatic Level Sensor User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 8
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 0
Safety Instructions
1. Prior to installation or removal:
Depressurize and vent system
Drain below sensor level
2. Con rm chemical compatibility before use.
3. Do not exceed maximum temperature/pressure speci cations.
4. Wear safety goggles or faceshield during installation/service.
5. Do not alter product construction.
6. Dispose of properly; DO NOT INCINERATE!
Signet 2250 Hydrostatic Level Sensor
1. Description
2. Speci cations
The Signet 2250 Hydrostatic Level Sensor uses a gauge pressure sensor to calculate the level of uid in a tank.
Gauge pressure sensors measure the difference in pressure between the process on one side of a diaphragm and the atmospheric
pressure on the opposite side of the diaphragm. A tiny capillary tube running from inside the sensor body and
up the length of the cable provides the reference to atmospheric pressure.
Hydrostatic level measurement can be a simple and practical solution to those applications where the uid
temperature is stable and constant.
IMPORTANT! This method of level measurement is not recommended if the uid is subject to
temperature variations or any other variable that causes the density of the uid to change.
The 2250 Hydrostatic Level Sensor has excellent chemical compatibility.
The wetted materials are PVDF, Ceramic and FPM.
Several different hardware mounting kits are available to accommodate every installation requirement.
3-2250.090 Rev. E 06/13 English
*3-2250.090*
English
Dimensions
½ in. (ANSI) or
25 mm (ISO)
Solvent Socket,
PVC-U
34 mm
(1.32 in)
115 mm
(4.55 in)
53 mm
(2.09 in)
General
Compatibility: Signet 8900 Multi-Parameter Controller
Signet 9900 Transmitter
Wetted Material
• Sensor housing: PVDF
• Union nut/bushing: PVC-U
• Diaphragm: Ceramic
Diaphragm seal: FPM (optional EPDM)
Cable type: 22 AWG, 3 conductor+capillary tube
Cable length: 10 m (32.8 ft) supplied
max. extension 120 m (400 ft)
Operating Temperature: 15 °C to 85 °C (5 °F to 185 °F)
Storage Temperature: -20 °C to 100 °C (-4 °F to 212 °F)
Operating Pressure: -XU: 0 to 0.7 bar (0 to 10 psig)
-XL: 0 to 3.4 bar (0 to 50 psig)
Electrical
Digital (S
3
L) Output:
Format: Serial ASCII, TTL level 9600 bps.
Accuracy: ±1% of full scale (±0.001 psi)
• Power: 5 VDC nominal, 1.5 mA max. current
(5 VDC min, 6.5 VDC max)
Reverse polarity and short circuit protected
4 to 20 mA Output:
• Factory span:
-XU: 0 to 0.7 bar (0 to 10 psig) =
0 m to 7 m (0 ft to 23 ft) of water
-XL: 0 to 3.4 bar (0 to 50 psig) =
0 m to 34 m (0 ft to 115.5 ft) of water
• Accuracy: ± 32 μA
• Resolution: < 5 μA
Response Time: < 100 ms
Loop Power Required: 12 to 24 VDC nominal, 22.1 mA max
(10.8 VDC min. to 26.4 VDC max.)
Max. Loop Impedance:
100 Ω @ 12 V
325 Ω @ 18 V
600 Ω @ 24 V
Standards and Approvals:
• CE
• RoHS Compliant
Manufactured under ISO 9001 for Quality,
ISO 14001 for Environmental Management and
OHSAS 18001 for Occupational Health and Safety.
China RoHS (Go to www.gfsignet.com for details)
This device complies with Part 15 of the FCC rules.
Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and,
(2) This device must accept any interference received,
including interference that may cause undesired operation.
Zobrazit stránku 0
1 2 3 4 5 6 7 8

Shrnutí obsahu

Strany 1 - *3-2250.090*

Safety Instructions 1. Prior to installation or removal:• Depressurize and vent system• Drain below sensor level 2. Confi rm chemical compatibili

Strany 2 - Digital (S

2 2250 Hydrostatic Level SensorInstallationIMPORTANT!A tiny capillary tube inside the sensor body is used to ensure that the back of the diaphragm rem

Strany 3 - Mounted through Tank Cover

32250 Hydrostatic Level SensorPLC3-2250-X3-9900-13-8052 3-2250-XX3-9900-13-8050 PLCNOCNCNO C NC ENTERENTER4-20 mA OutControl Relay Out 4-20 mA OutCont

Strany 4 - Control Relay Out

4 2250 Hydrostatic Level SensorThread the sensor cable through the O-ring, Union and through the ¾ in. conduit.Glue the union bushing onto the conduit

Strany 5 - I/O Module 3-8900.401-X

52250 Hydrostatic Level SensorDigital (S3L) wiring• 3-2250-11X-X options provide Digital (S3L) output when powered with 5 VDC from a 9900 Transmitter

Strany 6 - 4 to 20 mA Span

6 2250 Hydrostatic Level Sensor4 to 20 mA Loop Wiring• The 3-2250-21X-X models provide a 4 to 20 mA loop output when powered with 24 VDC.• Connectin

Strany 7 - DC 12 - 24 V

72250 Hydrostatic Level Sensor• The 4 to 20 mA span can be set to accommodate any application within the operating range of the sensor.• The respann

Strany 8 - 2250 Hydrostatic Level Sensor

Ordering InformationAccessories and Replacement Parts2250 Hydrostatic Level SensorMfr. Part No. Code Description5523-0322 159 000 761 Sensor cable (pe