Signet 8900 Multi-Parameter Controller
Multi-Parameter
Controller
C1 2.50 µS/cm
F2 58.43 GPM
Relay 1
Relay 2
ENTER
3-8900.090-1
3-8900.090-1 Rev. K 01/13 English
Contents
1. Speci cations 2
2. Description 4
2.1 Compatibility 4
2.2 System Overview 5
3. Installation and Basic Functionality 6
3.1 Unpacking 6
3.2 Tools and Equipment Required 6
4. Plug-in Modules 6
4.1 I/O Module 7
4.2 Power Module 7
4.3 Output Module 8
4.4 Relay Module 8
4.5 8059 External Relay Module 9
5. Mounting the Base Unit 10
5.1 Panel Mounting 10
5.2 Mounting Accessories 11
6. Wiring 12
6.1 Signal Type: Frequency 12
6.2 Digital (S
3
L) Input wiring 13
6.2.1 Digital Cable Length Calculations 14
6.2.2 Digital Cable Routing Diagrams 15
6.3 Power Module (3-8900.402-X) 17
6.4 Output Module (3-8900.405-X) 17
6.5 Relay Module (3-8900.403-X) 18
6.6 8059 External Relay Module 18
7. Operational Overview 19
7.1 General Operation Flowchart 19
7.2 System Con guration 20
7.3 Menu Directory 20
8. System Setup Menu 21
8.1 System Setup: Channel Type 21
8.2 Automatic Sensor Recognition 22
8.3 Automatic Channel Assignment 22
8.4 Automatic Display Con gurations 22
8.5 Other Automatic Displays 23
8.6 Displays for Derived Functions 23
Important Safety Information
Read and understand instruction manual before using
this product.
This unit is designed to be connected to equipment
which can be hazardous to persons and property if
used incorrectly. Read and understand all associated
equipment manuals and safety warnings before using
with this product.
Do not remove rear panel slot shields if the modules
are not installed.
Remove power to unit before wiring input and output
connections.
Wiring connections to this product should only be
performed by quali ed personnel.
Do not exceed published speci cations in the use of
this product.
In case of communication error or loss of source
signal:
Relays will be deactivated.
Current outputs will go to 22.1 mA.
Voltage outputs will go to 0 V.
9. Channel Settings 24
9.1 Channel Settings: Flow 24
9.2 Channel Settings: pH 25
9.3 Channel Settings: ORP 25
9.4 Channel Settings: Conductivity 26
9.5 Channel Settings: Pressure 27
9.6 Channel Settings: Temperature 27
9.7 Channel Settings: Level 28
9.8 Channel Settings: OTHER (4–20) 30
10. Hold Inputs 30
11. Relay Mode Descriptions 31
11.1 Relay Setup Menu 33
11.2 Advanced Relay Mode 34
12. Output 35
13. Calibration 36
13.1 Calibration: Flow 36
13.2 Calibration: pH 37
13.3 Calibration: ORP 37
13.4 Calibration: Conductivity 38
13.5 Calibration: Pressure 38
13.6 Calibration: Temperature 38
13.7 Calibration: Level 38
14. Options 39
14.1 Password Security 39
Appendix A: Derived Functions 40
Appendix B: Level System Con guration 41
Appendix C: Conductivity & Temp Comp 43
Appendix D: USP Limits 44
Appendix E: Cloning using Digital (S
3
L) 45
Appendix F: BTU Quick Start Guide 47
Appendix G: 8900 display messages 48
Ordering Information 52
Zobrazit stránku 0
1 2 3 4 5 6 ... 51 52

Shrnutí obsahu

Strany 1 - Controller

Signet 8900 Multi-Parameter ControllerMulti-ParameterControllerC1 2.50 µS/cmF2 58.43 GPMRelay 1Relay 2ENTER3-8900.090-13-8900.090-1 Rev. K 01/13

Strany 2 - IMPORTANT SAFETY INFORMATION

10Signet 8900 Multi-Parameter5. Mounting the Base UnitThe 8900 may be mounted in a panel, on a wall, or on virtually any surface including shelves, r

Strany 3 - 3 Signet 8900 Multi-Parameter

11 Signet 8900 Multi-Parameter5.2 Mounting AccessoriesMfr. Part No. Description3-8050.395 Splashproof rear cover • Use in conjunction with liq

Strany 4 - 2. Description

12Signet 8900 Multi-Parameter6. Wiring All wiring connections to the 8900 are made via the removable terminals of the plug-in modules. This section c

Strany 5

13 Signet 8900 Multi-Parameter6.2 Digital (S3L) Input wiring• The maximum cable length of the digital (S3L) bus varies depending on the types of sens

Strany 6

14Signet 8900 Multi-ParameterStep 1: Calculate the Total current requirements for digital (S3L) BranchesThis information will determine the total cur

Strany 7

15 Signet 8900 Multi-Parameter3-8900.621CI/O Module 3-8900.401-X1234567891011121314+5VDC (Black)Freq. Input (Red)GND (Shield)+5VDC (Black)Freq. Input

Strany 8 - 3-8900.620

16Signet 8900 Multi-Parameter3-8900.621CI/O Module 3-8900.401-X1234567891011121314+5VDC (Black)Freq. Input (Red)GND (Shield)+5VDC (Black)Freq. Input 2

Strany 9 - 8059-4AC

17 Signet 8900 Multi-Parameter6.3 Power Module (3-8900.402-X)• The 8900 can operate from a Universal AC power input (3-8900.402-1) or from a 24 VDC

Strany 10 - 5. Mounting the Base Unit

18Signet 8900 Multi-Parameter       l r er ale r er M6.5 Relay Module

Strany 11 - 1/2 NPT

19 Signet 8900 Multi-ParameterF1 123.45 GPMP3 4.32 pHF1 123.45 GPMF2 54.321 GPMRatio F1:F22.27P3 4.32 pHL4 19.49 cmRst1 1234567.8Gallons >Perm1 123

Strany 12 - 6. Wiring

2Signet 8900 Multi-ParameterGeneral Number of input channels: 2 or 4 or 6Input signal types:Digital (S3L): Serial ASCII, TTL level 9600 bps Freque

Strany 13 - Daisy-chain

20Signet 8900 Multi-Parameter7.3 Menu Directory• Select this item to exit View and go to the Menu Directory.• The Menu Directory may also be reach

Strany 14 - Maximum Cable (Metric)

21 Signet 8900 Multi-ParameterMenu Directory:System SetupReturning toViewMode...Channel 1 Type: None >Channel 1 Type:NoneChannel 1 Type:

Strany 15 - I/O Module 3-8900.401-X

22Signet 8900 Multi-ParameterpH ChannelThe default pH display shows the primary pH measurement on line 1 and the secondary temperature measurement on

Strany 16

23 Signet 8900 Multi-Parameter8.5 Other Automatic DisplaysThe 8900 will confi gure the following displays when appropriate:• For each channel assign

Strany 17 - 12 to 24 VDC

24Signet 8900 Multi-Parameter9.1 Channel Settings: If Channel Type = Flow: Set K-Factor (pulses per unit volume) from Flow Sensor manual:Minimum: 0.0

Strany 18 - AC or DC

25 Signet 8900 Multi-Parameter9.2 Channel Settings: If Channel Type = pH T1 Units: C >Select the unit of measure for the temperat

Strany 19 - 7. Operational Overview

26Signet 8900 Multi-ParameterSelect the unit of measure for this conductivity channel:uS/cm mS/cm k • cm M • cm PPM or PPB (of TDS)If un

Strany 20 - System Setup

27 Signet 8900 Multi-ParameterSelect the unit of measure for this pressure channel : psi bar kPa P1 Units: psi >Select decimal locati

Strany 21 - _ _ _ _ _ _ _ _

28Signet 8900 Multi-Parameter9.7 Channel Settings: If Channel Type = LevelSelect the unit of measure for this Level channel: cm m in ftYes: Le

Strany 22 - 1 Ext Relay(s)

29 Signet 8900 Multi-ParameterSelect the decimal location for the Volume display. Must be consistent with the range: XXXXX. XXXX.X XXX.XX

Strany 23 - Output2 22.10 mA

3 Signet 8900 Multi-ParameterAnalog Outputs (via I/O Modules and Output Modules)4 to 20 mA Output:Endpoints are adjustable and reversible:Minimum:

Strany 24 - **.*** >

30Signet 8900 Multi-ParameterHolds status of input, output and relay functions (fl ow totalizers will continue counting, however). Useful during calib

Strany 25 - Med >

31 Signet 8900 Multi-Parameter11. Relay Mode DescriptionsThe 8900 offers a wide selection of relay operating modes that enable the user to create com

Strany 26 - Signet 8900 Multi-Parameter

32Signet 8900 Multi-Parameter105ela utut ate0 t 00 PuleminenintPulse rate0 pulses100 pulsesStarting intProportional Pulsing• Proport

Strany 27 - Med >

33 Signet 8900 Multi-Parameter11.1 Relay Setup Menu Relay 1 Source: Flow F1 >Select measurement type or derived function to be assigned to

Strany 28

34Signet 8900 Multi-Parameter11.2 Multiple Relay Sources and Advanced Relay ModeWhen a Relay Source is set to "MULTIPLE" and the Relay Mode

Strany 29

35 Signet 8900 Multi-ParameterSelect measurement type or derived function to be assigned to this output.If I/O Module with Voltage Output is installed

Strany 30 - 5 10 30 60 ∞

36Signet 8900 Multi-Parameter13. CalibrationVolumetric calibrationSelect VOLUMETRIC CALIBRATION if the fl ow rate can be determined by fi lling a vessel

Strany 31 - Derived Functions

37 Signet 8900 Multi-ParameterSENSOR: Select to perform calibration at the sensor via 2750 EasyCal.INSTRUMENT: Select to perform calibration at th

Strany 32 - ( a &(b

38Signet 8900 Multi-Parameter13.4 Calibration: If Channel type = ConductivitySet Conductivity C1: >If C~ Cal at Instrument is selected,

Strany 33 - 5.0 GPM >

39 Signet 8900 Multi-ParameterAdjust LCD contrast: 4 settings.If two or more measurements of the same type are present, select a functional relations

Strany 34

4Signet 8900 Multi-Parameter2. DescriptionThe 8900 Multi-Parameter Controller takes the concept of modularity to the extreme. The controller comes wit

Strany 35 - 12. Output

40Signet 8900 Multi-Parameter• % Recovery is a derived measurement based on fl ow rate only. In a reverse osmosis system, % Recovery is fl ow rate in

Strany 36 - 13. Calibration

41 Signet 8900 Multi-ParameterZero reference point (Z):Defi nition: The point in the vessel where you want the 8900 to display zero (0 ft, 0 gal. etc)

Strany 37 - >

42Signet 8900 Multi-ParameterCustom point #103000 gal.Custom point #10 gal.Custom point #245 gal.Custom point #3118 gal.Custom point #4Custom point #5

Strany 38 - Factory Cal: >

43 Signet 8900 Multi-ParameterTDS FactorSome industries need to display a conductivity value as Total Dissolved Solids (TDS), measured in units of par

Strany 39

44Signet 8900 Multi-ParameterAppendix D: USP LimitsUSP (United States Pharmacopoeia) has defi ned a set of conductivity values (limits) to be used for

Strany 40 - RO Membrane

45 Signet 8900 Multi-ParameterAppendix E: Cloning using Digital (S3L) (Serial Cloning)Cloning enables copying the programming from one unit into

Strany 41 - O(ffset)  0

46Signet 8900 Multi-ParameterStep 4: Confi gure the Transmitting Controller1. Now go to the Transmitting Controller’s Option menu and scroll until yo

Strany 42 - • SG • V

47 Signet 8900 Multi-ParameterAppendix F: BTU Quick Start GuideThis guide outlines how to confi gure the 8900 Multi-Parameter to calculate Power and E

Strany 43

48Signet 8900 Multi-ParameterAppendix G: 8900 display messages8900Multi-ParameterSavingReturning toMenu DirectoryReturning toViewMode...Enter Passwor

Strany 44 - Appendix D: USP Limits

49 Signet 8900 Multi-ParameterChannel 1 ID:Not Assigned?Input ModuleNot FoundInput/Output 1&2Module DefectiveComm/Out 3&4Module DefectiveCal P

Strany 45 - (Serial Cloning)

5 Signet 8900 Multi-Parameter2.2 System OverviewThe most basic 8900 system consists of a Base Unit, an I/O Module and a Power Module. Outputs and rel

Strany 46

50Signet 8900 Multi-ParameterNotes

Strany 47 - T3 Units:

51 Signet 8900 Multi-ParameterNotes

Strany 48

Ordering InformationMfr. Part No. Code DescriptionBase Unit, required; choose one3-8900 159 000 868 Base unit with back-lit LCDI/O Module,

Strany 49

6Signet 8900 Multi-Parameter3. Installation and Basic FunctionalityThis section lists all items included with individual 8900 system components, descr

Strany 50

7 Signet 8900 Multi-Parameter4.1 I/O Module, required; install one (1) Mfr. Part No. Description3-8900.401-1 Two inputs, no outputs3-8900.401-

Strany 51

8Signet 8900 Multi-Parameter4.3 Output Module, optional; install zero (0) or one (1)Mfr. Part No. Description3-8900.405-1 Two passive 4 to 20

Strany 52 - Spare Parts

9 Signet 8900 Multi-Parameter8059-4NC COM NORELAY ANC COM NORELAY BNC COM NORELAY CNC COM NORELAY DDC OUTPUT+ -DC INPUT- + + S3L -+